TGA Action ID RC-2026-RN-00719-1
Type of Product Medical Device
Product Name/Description Arjo Carevo Shower Trolley

Model number: BAC1113-01

Date of manufacture: 01-Aug-2015 and 29-Oct-2025

Multiple serial numbers
ARTG Number 283231
Action Type Product Correction
Action Level Consumer
Hazard Classification Class II
Reason for the Action Arjo has received 10 reports of events, including 7 injuries, where the side support opened unintentionally during use. Internal investigation showed that the side support of the Carevo shower trolley may open unintentionally when the handles are not fully locked and additional force is applied during patient handling. This can lead to a resident/patient fall and serious injury, such as head injury, fractures, or other trauma resulting from a patient fall.

Two factors have been identified which may lead to unintentional side support opening:

1. Incorrect or incomplete handle locking combined with user movement during handling. To support correct operation and reduce the likelihood of unintentional side support opening, the Instructions for Use have been updated to reinforce proper locking of both handles and the recommended ergonomic working posture.

2. For a subset of devices, the torsion springs may not always return the handle to its fully locked position. This may result in the handle remaining partially engaged, leading to unintentional side support opening if one handle is accidentally activated.
Instructions for Customers/Users Customers are to:
· Review the additional warnings within the updated Instructions for Use (IFU).
· Use the device in accordance with the updated IFU.

For devices manufactured between 21-Feb-2024 and 29-Oct-2025, Arjo will contact customers to arrange an on-site visit to replace the torsion springs.
Action Commencement Date 28/09/2026
Responsible Entity Arjo Australia Pty Ltd
Contact Information 1800 002 756 - Arjo Australia