Print version of this report
TGA Action ID RC-2026-RN-00717-1
Type of Product Medical Device
Product Name/Description Certain Philips MR Systems with Diffusion Tensor Imaging (DTI)

Software Versions: R11, R12, & R13

Ingenia 1.5T
Models: 781341, 781396

Ingenia 3.0T
Models: 781342, 781377

Ingenia 3.0T CX
Model: 781271

Ingenia Ambition 1.5T X
Model: 781356

Ingenia Elition S 3.0T
Model: 781357

Ingenia Elition X 3.0T
Model: 781358

Ingenia Elition S
Model: 782106

MR 7700
Models: 782120, 782155
ARTG Number 230064, 372370, 485786
Unique Device Identifier (UDI) N/A
Action Type Product Correction
Action Level Hospital
Hazard Classification Class II
Reason for the Action When exporting diffusion tensor imaging (DTI) scans from a Philips MR system using the Enhanced MR DICOM format, the exports contain incorrect data (diffusion directions). If the Enhanced MR DICOM export is used in conjunction with third-party software for tractography (e.g. for mapping nerve fiber pathways in the brain), the software will not be able to correctly process the DTI data for reconstruction.

If an incorrect reconstruction image is used for treatment planning, it may lead to inadequate or inappropriate treatment. As of August 2026, there have been 2 complaints associated with this problem, with no reports of injury or adverse events.
Instructions for Customers/Users Philips will contact customers to schedule a software upgrade to fix the problem.

In the interim, customers may continue to use their systems in accordance with the intended use. To prevent the problem from occurring, customers will need to use the Classic DICOM transfer option for exporting DTI scans, per the instructions in the Customer Letter.
Action Commencement Date 21/09/2026
Responsible Entity Philips Electronics Australia Ltd
Contact Information 1800 251 400 - Philips Service Delivery Team