Print version of this report
TGA Action ID RC-2026-RN-00696-1
Type of Product Medical Device
Product Name/Description Newport HT70, Newport HT70 Plus Ventilators and PCBA Service Parts

Multiple Model Numbers

Multiple GTINs

Multiple Serial Numbers

ARTG 142414 (cancelled 30/10/2024)
ARTG Number
Unique Device Identifier (UDI) 10884521542266
Action Type Recall
Action Level Hospital
Hazard Classification Class I
Reason for the Action This is an update to the previously closed market action, TGA reference RC-2025-RN-00483-1.
Medtronic identified a limited subset of potentially affected goods for which the end consignee could not be determined due to the age of the products and limitations in historical distribution records. Out of an abundance of caution, all known HT70 consignees will be notified regarding these goods.

Investigation into customer complaints have identified two separate capacitors on one of the ventilator’s Printed Circuit Board Assemblies (PCBAs), that, in case of failure, may result in one of two distinct outcomes: The ventilator either shuts down during use, thus necessitating the use of an alternate form of ventilation, or the shutdown alert alarm fails to alarm effectively during shut down.

Cessation of operation and ventilation or loss of audible shutdown alarm in a mechanically ventilated patient may result in respiratory failure, hypoventilation, low oxygen saturation, hypoxia, delay to treatment, or potentially severe consequences like brain injury and/or death.

Medtronic has identified 73 complaints and adverse health events associated with these problems, including one serious injury and two deaths.
Instructions for Customers/Users The customer must remove all affected devices from service and replace with an alternate means of ventilation.

Healthcare providers with patients currently using the ventilators should assess the overall risk and appropriate timing for replacement of the devices.

After completing the transition to alternate ventilator(s), the customer should quarantine all devices and service parts and follow their facility’s procedures for removal and disposal.
Action Commencement Date 17/09/2026
Responsible Entity Medtronic Australasia Pty Ltd
Contact Information rs.anzpmv@medtronic.com - Medtronic Australasia