TGA Action ID RC-2026-RN-00682-1
Type of Product Medical Device
Product Name/Description Philips CT Rembra systems

Reference: 728381

Software version 6.0.0.X
ARTG Number 352131
Unique Device Identifier (UDI) N/A
Action Type Product Correction
Action Level Hospital
Hazard Classification Class II
Reason for the Action Philips has identified multiple software problems in CT Rembra software version 6.0.0.X.

Problems:
1. 6.0.0.X - ECG-gated cardiac axial scan image reconstruction failure
2. 6.0.0.X - Distance between inner and outer lasers changes when gantry is tilted
3. 6.0.0.X - Tracker aborts with auto voice setting
4. 6.0.0.14420 - Reconstruction failure with STD-QA exam card
5. 6.0.0.X - Bad detectors cannot be automatically added into the Bad Detector Table
6. 6.0.0.X - Clinical scan startup not prohibited with insufficient tube heat capacity
7. 6.0.0.X - CDM board may present corrupted data:

These problems may result in reconstruction failures, scan interruptions, image artifacts, positioning inaccuracies, or corrupted images.
To date there have been no injuries /rescan associated with these problems.
Instructions for Customers/Users Customers may continue to use their system(s) in accordance with the guidance provided in the customer letter. Philips has identified the affected installed base and will contact you to provide a software update to resolve these problems (reference FCO72800855).
Action Commencement Date 7/09/2026
Responsible Entity Philips Electronics Australia Ltd
Contact Information 1800 251 400 - Philips Service Delivery Team