TGA Action ID RC-2026-RN-00657-1
Type of Product Medical Device
Product Name/Description Philips iCT, Ingenuity CT, and Brilliance CT 64 systems

Software Version: 4.1.10.x

Product Codes: 728306, 728321, 728323, 728326
ARTG Number 321687, 423851, 445069
Action Type Product Correction
Action Level Hospital
Hazard Classification Class II
Reason for the Action During the cardiac scan planning, if the X (XCA, XCB, XCC, XCD) filter for reconstruction is selected, the reconstructed cardiac images may display unexpected noise.

Noisy cardiac images may result in reduced image quality, including recognisable artifacts, anatomy obscured by artifacts, or non-diagnostic images.

To date Philips has not received any reports of injury or patient harm associated with this problem
Instructions for Customers/Users Customers may continue to use system(s) in accordance with the intended use. Users may use alternative CA, CB, CC, and CD filters and evaluate the image quality during live planning or during offline reconstruction.

Philips will contact customers to schedule a time for a Philips Field Service Engineer (FSE) to visit affected sites and install a software solution (Reference FCO72800857) to resolve the problem.
Action Commencement Date 28/08/2026
Responsible Entity Philips Electronics Australia Ltd
Contact Information 1800 251 400 - Philips Service Delivery Team