TGA Action ID RC-2026-RN-00656-1
Type of Product Medical Device
Product Name/Description LEO+ Self-Expanding Nitinol Stent (Stent, vascular, intracranial) and Eclipse Balloon Catheter (Catheter, intravascular, occluding)

Product Codes:LEO55X50, LEO45X40, LEO45X40, LEO55X75, LEO40X30, LEO45X30, LEO45X25, LEO40X35, LEO45X25, LEO45X30, SECLIPSE7

Lots: 633957, 638339,639251, 639961, 643012, 643348, 645061, 646869, 648969, 649236
ARTG Number 155085, 170160
Unique Device Identifier (UDI) 03700481335338, 03700481335215, 03700481337912, 03700481330272, 03700481335192
Action Type Recall
Action Level Hospital
Hazard Classification Class II
Reason for the Action Routine bacterial endotoxin quality control testing performed by the manufacturer, Balt Extrusion SAS (France), returned non-conforming results on sampled ANRYV connectors.

Two production lots of the ANRYV connector did not meet the established endotoxin acceptance criteria. As testing is performed on a sampling basis, the presence of non-conforming devices within the affected lots cannot be excluded.

Elevated endotoxin levels in a device that contacts the bloodstream may cause a pyrogenic reaction (e.g. fever or inflammatory response) in the patient. No patient injuries related to this problem have been reported.

The LEO+ stent and Eclipse balloon catheter devices themselves are not affected by this problem; however, as the connector lot cannot be identified without opening the packaging, the complete units are being recalled.
Instructions for Customers/Users Customers should inspect their stock and quarantine any affected units of the lots listed. A Balt Australia representative will arrange collection of the affected units and delivery of replacement stock
Action Commencement Date 3/09/2026
Responsible Entity Balt Australia Pty Ltd
Contact Information 07 2111 7995 - Balt Australia Pty Ltd Customer Service