TGA Action ID RC-2026-RN-00651-1
Type of Product Medical Device
Product Name/Description VITROS Immunodiagnostic Products Anti-HAV Total Reagent Pack and Total Calibrator. An in vitro diagnostic medical device (IVD)

Catalogue Numbers: 6801461 and 6801462

Affected Lots: 6672, 6680, 6690, 6720, 6740, 6760

Expiry Dates: 15 August 2026, 15 September 2026, 10 October 2026, 1 December 2026, 5 January 2027, 27 January 2027
ARTG Number 266611
Action Type Product Correction
Action Level Hospital
Hazard Classification Class II
Reason for the Action QuidelOrtho has confirmed that certain VITROS Immunodiagnostic Products Anti-HAV Total Reagent Packs and Total Calibrators may experience calibration or quality control (QC) failures.

QC may be affected in the following ways:
- For positive controls, this may cause results to shift to Borderline or Negative.
- For negative controls, this may cause results to shift to Borderline or Positive

Patient samples may be affected; however, investigation has not observed an impact on patient samples
Instructions for Customers/Users It is acceptable to continue using an affected lot of VITROS Anti-HAV Total Reagent Pack, provided your laboratory has successfully calibrated and generated acceptable QC results using the affected lot.

If your laboratory has experienced calibration and/or QC failures, discontinue using, render unusable, and discard the affected lot(s) of VITROS Anti-HAV Total Reagent Pack and associated lot of VITROS Anti-HAV Total Calibrator. QuidelOrtho will credit or replace your discarded inventory

QuidelOrtho does not recommend a review of previously reported results, as our investigation has not generated evidence that patient samples are affected by this problem.

The results of any diagnostic test should be evaluated in conjunction with a patient’s history, risk factors, clinical presentations, signs, and symptoms, as well as the results of other tests.
Action Commencement Date 27/08/2026
Responsible Entity Ortho-Clinical Diagnostics Australia Pty Ltd
Contact Information 1800 032 359 - Technical Services Center or mikaela.rasile@quidelortho.com