| TGA Action ID |
RC-2026-RN-00649-1 |
| Type of Product |
Medical Device |
| Product Name/Description |
Dexcom G7 iOS App
SW12300 G7 App (iOS) Version: 2.12 |
| ARTG Number |
529343 |
| Action Type |
Product Correction |
| Action Level |
Consumer |
| Hazard Classification |
Class II |
| Reason for the Action |
In G7 iOS App version 2.12, only Critical Alerts are designed to override the phone’s Silent Mode or Focus Mode setting. Non-critical alerts behave in accordance with the phone’s notification settings. In the affected version of the Dexcom G7 iOS app, non-critical alerts may not sound if the phone is in Silent Mode or Focus Mode.
If a user is unaware that non-critical alerts may not sound due to behaving in accordance with the phone settings, there is a potential for delayed detection of a hypoglycaemic or hyperglycaemic event. |
| Instructions for Customers/Users |
For active users of the Dexcom G7 iOS App, please complete the following actions: 1. Open the App Store and search for the Dexcom G7 iOS App. 2. Install the latest version of the Dexcom G7 iOS App.
This action has been closed out on 25/11/2026 |
| Action Commencement Date |
24/08/2026 |
| Responsible Entity |
|
| Contact Information |
1300 851 056 - Dexcom Technical Support |