TGA Action ID RC-2026-RN-00647-1
Type of Product Medical Device
Product Name/Description MEDUMAT Standard2 ventilators

Product codes: WM 28710-01, WM 28710-02, WM 28710-03, WM 28710-04

Sales variants: WM 29300, WM 29500, WM 29350, WM 29550

Serial numbers: 14756, 17247, 17248 and 17249
ARTG Number 221469
Unique Device Identifier (UDI) 010405465234805
Action Type Product Correction
Action Level Hospital
Hazard Classification Class II
Reason for the Action Post-market surveillance by the manufacturer identified usage-related incidents in emergency ventilation caused by undetected leakage and disconnections in the breathing circuit (mask, airway device, ventilation hose, measuring circuit, PEEP control tube, pressure measuring tube, Flow Check sensor, CO2 measuring tube, elbow, breathing system filter).

These leaks/disconnections may prevent effective ventilation and expose patients to life-threatening injuries.

Between 2023 and 2026, nine (9) incidents have been reported, including three (3) reports of serious injury.
Instructions for Customers/Users Customers are to:
- Display the 'Information on undetected leakage during ventilation with MEDUMAT Standard2' notice in relevant operational areas.
- Review and implement the supplement to the instructions for use. Perform pre-use function checks and monitoring for leaks/disconnections and responding appropriately to alarms such as “Airway pressure low”, “Apnea”, “MVe low”, and “Leak on patient side.”
- Ensure that all operators are trained in the recognition of measured values indicating leakage (low/no Vte, low/no MVe, insufficient pressure build-up).
Action Commencement Date 26/08/2026
Responsible Entity Rapp Australia Pty Ltd
Contact Information 0418 108 834 - Roger Buckle, RAPP Australia