TGA Action ID RC-2026-RN-00646-1
Type of Product Medical Device
Product Name/Description Carescape Central Station (CSCS) V3.0.x

CSCS V3 MAS800 Desktop
Product Code: STU
Part Numbers: 5867474-04, 5867474-13
GTIN: 00195278343512, 00195278516367

CSCS V3 MAI800 Integrated
Product Code: STV
Part Numbers: 5867474-03
GTIN: 00195278343383
ARTG Number 132673
Action Type Product Correction
Action Level Hospital
Hazard Classification Class II
Reason for the Action GE HealthCare has become aware that the Carescape Central Station (CSCS) V3.0.x may not resume patient monitoring for at least 3 minutes following a system restart. This situation can arise only when the ‘Allow list database is corrupted’ message is displayed in the title bar of the CSCS prior to a system restart. In the unlikely event that this situation occurs, it can lead to a potential delay in recognition of clinically significant events.

There have been no injuries reported to GE HealthCare as a result of this problem.
Instructions for Customers/Users - Customers are to review the actions required for continued use of the device in the Customer Letter.
- GE HealthCare will provide a User Manual addendum that provides guidance regarding the prolonged startup condition.
Action Commencement Date 24/08/2026
Responsible Entity GE Healthcare Australia Pty Ltd
Contact Information 1800 659 465 - GE Healthcare