TGA Action ID RC-2026-RN-00642-1
Type of Product Medical Device
Product Name/Description Imager II Angiographic Catheters

All Lot Numbers

Multiple Inner and Outer Package UPNs

Expiration Dates: 08/10/2026 - 09/06/2028
ARTG Number 433411
Action Type Recall
Action Level Hospital
Hazard Classification Class II
Reason for the Action Boston Scientific is initiating an urgent recall of Imager II Angiographic Catheters due to a manufacturing problem affecting the catheter tip. Boston Scientific has determined that these Catheters were manufactured with reduced levels of stabilising agents in the tips. This may increase the risk of tip degradation and detachment.

Although no increase in tip detachment complaints has been identified for the affected lots, the devices were manufactured outside the approved design specification and are being recalled from the market.

The most common adverse health consequence of tip detachment in the intravascular space is a delay in procedure to exchange for a new device. The most serious potential adverse health consequence within the intravascular space is life threatening embolism resulting from a device fragment obstructing blood flow resulting in end organ failure.
Instructions for Customers/Users Customers are to:
- Immediately stop further use or distribution and quarantine the affected product.
- Return the affected product to Boston Scientific in accordance with the instructions in the Customer Letter.
- Continue to follow the standard of care when monitoring patients treated with the product.
Action Commencement Date 21/08/2026
Responsible Entity Boston Scientific Pty Ltd