Print version of this report
TGA Action ID RC-2026-RN-00622-1
Type of Product Medical Device
Product Name/Description VITROS Chemistry Products VALP Reagent. An in vitro diagnostic medical device (IVD).

Catalogue Number: 680 1710

Lots: 2511-35-3174, 2511-36-3272, 2511-37-3554

Expiry: 17-Aug-2026, 01-Nov-2026, 27-Feb-2027
ARTG Number 245278
Action Type Recall
Action Level Hospital
Hazard Classification Class II
Reason for the Action QuidelOrtho confirmed a problem involving VITROS Chemistry Products VALP Reagent. Lots from Generations (GENs) 35, 36, and 37 may generate positively biased results. QuidelOrtho’s investigation into the problem determined that lots from GENs 35, 36, and 37 of VITROS VALP Reagent may experience a drift in performance across the shelf life of the product. A drift in performance may cause positively biased results as the product ages.

To date, no adverse events have been reported due to this problem.
Instructions for Customers/Users Customers are to:
• Discontinue using, render unusable, and discard all inventory of affected lots of VITROS VALP Reagent.
• Complete the Customer Response Form. QuidelOrtho will credit or replace discarded inventory.
Action Commencement Date 14/08/2026
Responsible Entity Ortho-Clinical Diagnostics Australia Pty Ltd
Contact Information 1800 032 359 - QuidelOrtho Technical Services