Print version of this report
TGA Action ID RC-2026-RN-00602-1
Type of Product Medical Device
Product Name/Description Getinge Assured Batch Monitoring PCD

Model Number: PL589/ 6005500575

Serial Number: 54043A

Manufacture Dates: 04 Nov 2025 – 27 Mar 2026
ARTG Number 227239
Action Type Recall
Action Level Hospital
Hazard Classification Class III
Reason for the Action Getinge has identified a problem with certain Getinge Assured Batch Monitoring PCD units, where the required O-ring was not installed during manufacture following a customer complaint.

The root cause was identified as a combination of a supplier production process failure and an internal quality inspection error.

Failure of the device, due to the missing O-ring, may state sterilisation cycle was not completed and may require re-sterilisation, which may disrupt workflow. This could result in delay of goods and, as such, delay of procedure; should this occur, it is anticipated that any procedural delays would be prior to administration of anaesthesia.
Instructions for Customers/Users Customers are to:
- Examine their inventory immediately to identify any affected Getinge Assured Batch Monitoring PCD units listed in this Product Recall and remove them from use.
- Arrange return of the affected product by emailing a completed response form to Quality.AUNZ@getinge.com.
Action Commencement Date 12/08/2026
Responsible Entity Getinge Australia Pty Ltd
Contact Information 1800 438 464 - Getinge Australia