TGA Action ID RC-2026-RN-00594-1
Type of Product Medical Device
Product Name/Description StrykeFlow PS and AHTO Tube Sets

Strykeflow PS Suction Irrigator Single Spike Tube Set
Catalogue Number: 0250070530
Lot Numbers: 24260FG2, 24261FG2, 24298FG2, 24299FG2, 24302FG2, 24341FG2, 25008FG2, 25010FG2, 25011FG2, 25024FG2, 25034FG2, 25092FG2, 25182FG2, 25183FG2, 25216FG2, 24297FG2, 25035FG2, 25063FG2, 25064FG2, 25093FG2

AHTO Suction Irrigation Dual Spike Tube Set
Catalogue Number: 0250070640
Lot Numbers: 24289FG2, 24338FG2, 25009FG2, 25091FG2
ARTG Number 207650
Action Type Recall
Action Level Hospital
Hazard Classification Class II
Reason for the Action There has been an increase in reported complaints describing leaking from the device during use.

During use, the device may experience fluid leakage from the handpiece, which may result in fluid accumulating on the floor. For the AHTO tube sets, saline may react with internal copper wiring, resulting in the release of vaporised saline that may resemble smoke.

The potential impacts include user inconvenience, short procedural delays, skin, eye or mucosal membrane irritation, contamination of the surgical field, perceptions of burning, and an increased risk of slip and fall incidents.
Instructions for Customers/Users Customers are to:
1. Quarantine any affected product.
2. Contact their local Stryker representative to coordinate the return and replacement of any affected product.
Action Commencement Date 7/08/2026
Responsible Entity Stryker Australia Pty Ltd