TGA Action ID RC-2026-RN-00586-1
Type of Product Medical Device
Product Name/Description Adaptable Beach Chair Positioning Kit

Product code: 804-33-300

Lot: 2023032001
ARTG Number 396179
Action Type Recall
Action Level Wholesale
Hazard Classification Class I
Reason for the Action Quantum OPS Adaptable Beach Chair Positioning Kits may be affected by a fracture of a welded joint below the patient headrest. The affected welded joint provides structural support for the patient headrest. The potential consequences include loss of head support, unintended patient movement, disruption to airway management, interruption to the surgical procedure and/or disruption to surgical workflow.

Fracture of the welded joint supporting the patient headrest may result in partial or complete loss of head support and/or unintended movement of the headrest during a surgical procedure. This may result in patient movement, disruption to airway management, interruption or delay to the surgical procedure, and, in the worst case, patient injury associated with loss of head support or compromised airway management.

This recall does not affect any other Quantum Ops products.
Instructions for Customers/Users The affected customer should return stock per the instructions in the Customer Letter.

No alternative stock is available currently. Alternative stock is expected to be available from mid-October and will be available for order.

This action has been closed out on 16/11/2026
Action Commencement Date 12/08/2026
Responsible Entity Enovis Surgical Australia Pty Ltd
Contact Information 1300 667 730 - Enovis Surgical Australia Customer Service