Print version of this report
TGA Action ID RC-2026-RN-00581-1
Type of Product Medical Device
Product Name/Description GS1-HT Extraction Script for use within EasyScreen Sample Processing Kit Workflow. An in vitro diagnostic medical device (IVD).

Multiple Computer Names

Multiple Computer Serial Numbers

Script Versions: SP012M_ERS_v1.0.0, SP012HT_ERS_v1.7.8
ARTG Number 224568
Action Type Product Correction
Action Level Hospital
Hazard Classification Class II
Reason for the Action The impacted extraction scripts contain a missing section of error-handling code for sample positions 81–95. If a low-volume or missing-sample condition occurs in positions 81 - 95, the script may apply recovery logic intended for positions 1 - 80. This can cause the liquid handler to move to preceding specimen positions and aspirate material from an incorrect tube into the corresponding deep-well plate.

To date, no adverse events have been reported due to this problem.
Instructions for Customers/Users Customers are to:
• Follow the interim measures outlined in the Customer Letter for continued use of the product prior to installation of the correction.
• Genetic Signatures has amended the extraction-script error-handling logic for sample positions 81–95.
• Authorised personnel will install the corrected script and verify and record the installed version before the systems are returned to routine use.
• No EasyScreen Sample Processing Kit reagents or consumables need to be quarantined or returned because the physical kit components are not affected.

This action has been closed out on 05/11/2026
Action Commencement Date 3/08/2026
Responsible Entity Genetic Signatures Ltd
Contact Information +353 877734444 - Genetic Signatures Ltd