TGA Action ID RC-2026-RN-00559-1
Type of Product Medical Device
Product Name/Description CombiDiagnost R90, DigitalDiagnost 4, DigitalDiagnost C90, and ProxiDiagnost N90

CombiDiagnost R90
Model: 709031
Version: 1.1.7V08.4270.006

DigitalDiagnost 4 Flex / Value
Model: 712028
Version: 4.3.2V18.4270.005

DigitalDiagnost 4 High Performance
Model: 712031
Version: 4.3.2V18.4270.005

DigitalDiagnost C90 High Performance
Model: 712034
Version: 1.1.7V20.4270.008
ARTG Number 117662, 340852, 390959
Action Type Product Correction
Action Level Hospital
Hazard Classification Class II
Reason for the Action Following implementation of PSC 10.1, Philips identified several problems affecting DigitalDiagnost and ProxiDiagnost/ CombiDiagnost systems:

1. The system may not enter a ready state for wall stand exposures when the Ceiling Suspension Motorised (CSM) Arm Swing is not in the locked position, preventing certain examinations from being performed as intended.
2. The Tube Control Handle (TCH) may intermittently restart multiple times during start-up, resulting in delays before the system is available for clinical use.
3. Systems without a valid 2D camera licence may display the live camera image on the Acquired Work Station (AWS).
4. Following a language change, the TCH display may not immediately synchronise with the selected system language after restart.
5. The CSHeightOffset calculation used by the Smart Collimation Thorax (SCT) feature may be less accurate for source-to-image distances greater than 150 cm.

These problems may result in temporary delays, limited functionality, or display inconsistencies; however, clinical examinations can continue using alternative workflows where applicable. Philips will provide a software update to address these problems.
Instructions for Customers/Users Customers are to:
- Continue using the affected systems in accordance with the intended use until the software update is installed.
- Note that the final correction, consisting of a software update and physical kit, is planned for release from 31 August 2026.
Action Commencement Date 6/08/2026
Responsible Entity Philips Electronics Australia Ltd
Contact Information 1800 251 400 - Philips Service Delivery Team