TGA Action ID RC-2026-RN-00556-1
Type of Product Medical Device
Product Name/Description VITEK 2 System. An in vitro diagnostic medical device (IVD).

Software versions before V11.0
ARTG Number 287987
Action Type Product Correction
Action Level Hospital
Hazard Classification Class II
Reason for the Action In 2022, during a historical review of VITEK 2 System software BIOART rules, bioMérieux identified four (4) discrepancies that could potentially affect the interpretation and reporting of antimicrobial susceptibility testing (AST) results in Australia. This was communicated in RC-2022-RN-00640-1.

This action is to provide a permanent corrective action for the identified BIOART rule discrepancies.
Instructions for Customers/Users As a permanent corrective action, bioMérieux has corrected the identified BIOART rule discrepancies in VITEK 2 Software Version 11.0.

bioMérieux team will contact customers in due course to arrange a mutually convenient time to update VITEK 2 Systems to Software Version 11.0.
Action Commencement Date 23/07/2026
Responsible Entity Biomerieux Australia Pty Ltd
Contact Information 1800 333 421 - Helpdesk