| TGA Action ID |
RC-2026-RN-00544-1 |
| Type of Product |
Medical Device |
| Product Name/Description |
Packs containing BD Connecta stopcock
Product Descriptions: The Mount Pacemaker Pack; Flinders Central Line Pack; St Radiofrequency Pack; Radiofrequency Ablation; Infant Lumbar Pack; St Universal Lumbar Puncture Kit; Adult Lumbar – Sterile Items; St Lumbar Puncture Kit 2; St Flexi Cysto Pack
Product Codes: 08-370, 23-925, 25-808, 25-845, 26-099, 26-108, 26-188, 37-925, PR37-759
Batch Numbers: Multiple |
| ARTG Number |
163168, 163252, 164294 |
| Action Type |
Product Alert |
| Action Level |
Hospital |
| Hazard Classification |
Class II |
| Reason for the Action |
The listed packs contain the BD Connecta stopcock supplied by Becton Dickinson Pty Ltd (BD), which was the subject of a Product Alert (TGA ref: RC-2026-RN-00446-1).
BD has identified an observed problem in how Port C of the BD Connecta Stopcocks operates during use. Reports received by BD indicate that when using polypropylene plastic luer lock syringes, the connection at Port C may over thread the syringe and continue to spin. This behaviour has led some clinical users to perceive the connection as unstable.
Although BD testing has demonstrated that a secure connection can be achieved at Port C even in instances of continued rotation (over-threading), the differing torque and tactile resistance required for Port C compared to other ports may lead users to conclude that additional corrective steps are necessary and apply unnecessary manipulation, repositioning, or force to the connection.
Under such circumstances, connection integrity could be adversely affected, which may result in interruption or delay of therapy and, in some cases, leakage at the connection site. Such outcomes could lead to under delivery of fluids or medications. |
| Instructions for Customers/Users |
Customers should: 1. Review the Customer Letter and take the actions as described below. • As per current Instructions for Use: o Ensure all connections are secure before use and verify that injection ports are closed before and after each use. o When using the product, avoid over-threading. Excessive threading may damage the integrity of the product. o Use of excessive force during connection can over stress stopcock ports.
2. Limit the use of polypropylene plastic luer lock syringes in port C, or if needed be mindful of the problem noted above. |
| Action Commencement Date |
21/07/2026 |
| Responsible Entity |
|
| Contact Information |
(02) 8718 2888 - Multigate Medical Products |