Print version of this report
TGA Action ID RC-2026-RN-00541-1
Type of Product Medical Device
Product Name/Description Silver Graft Vascular Prosthesis

Product Descriptions: Silver Graft Straight Tube 06MM 60CM and Silver Graft Bifurcation 16X8MM 40CM

Product Codes: 1108012 & 1108015

Batches: 2240914076 & 2234723914

Supplied via the Special Access Scheme (SAS)
ARTG Number
Action Type Product Alert
Action Level Hospital
Hazard Classification Class III
Reason for the Action The products are currently labelled as "MR Safe" on several labelling components, including the primary label, secondary label, Instructions for Use (IFU), and implant card. However, as these medical devices contain metallic components, "MR Conditional" is considered the appropriate MR safety classification in accordance with ASTM F2503.

However, the medical device specific MRI conditions are correctly described in the valid Instructions for Use (Doc. ID 12158031) and in the patient information documentation. The MRI assessment and medical device verifications have confirmed that the affected medical devices can be safely used under the specified MRI conditions.
Instructions for Customers/Users Customers are to confirm understanding of the relevant information within the Patient Information and valid IFU.

This includes that Magnetic resonance Imaging (MRI) can be safely done in patients with implanted Silver Graft vascular prostheses under the following conditions:
- Static magnetic field: 1.5 T and 3 T
- Maximum spatial field gradient: 6.6 T/m
- Whole body averaged specific absorption rate (SAR): 2 W/kg

It is recommended that this information is provided to patients so they are aware of the MR Conditional nature of the implant, follow the guidance in the Patient Information document, and appropriately advise their treating clinician of the implant prior to an MRI examination.

This action has been closed out on 1/09/2026
Action Commencement Date 21/07/2026
Responsible Entity B Braun Australia Pty Ltd
Contact Information 1800 251 705 - B Braun Customer Service