TGA Action ID RC-2026-RN-00539-1
Type of Product Medical Device
Product Name/Description Philips Allura R8.2 Systems, equipped with any AD7x patient table

Allura Xper FD10
Model: 722026

Allura Xper FD10/10
Model: 722027

Allura Xper FD20
Model: 722028

Allura Xper FD20/15
Model: 722058
ARTG Number 225815
Action Type Product Correction
Action Level Hospital
Hazard Classification Class II
Reason for the Action In some cases, following a system (re)start, manual and motorised geometry movements may be disabled due to a firmware problem in the Advanced Motion Controller (AMC) single-axis motion control unit used in the AD7x tables of Allura R8.2 systems. When the problem occurs, no message is displayed to the user. In this situation, X-ray imaging is available. The problem can be resolved by performing a cold restart of the system or by restarting the geometry.

Philips has not received reports of harm associated with this problem on the devices in scope of this correction.
Instructions for Customers/Users If the problem occurs, customers are to restart the geometry following the steps described in the Instructions for Use and customer letter. A geometry restart may take up to 2 minutes to complete.

Philips will correct affected systems by implementing a software update (R8.2.102.1). Philips currently expects R8.2.102.1 to be available by Q1 2027, subject to regulatory clearance.

Your local Philips representative will contact you to schedule a visit for the implementation of the software update once it becomes available.
Action Commencement Date 22/07/2026
Responsible Entity Philips Electronics Australia Ltd
Contact Information 1800 251 400 - Philips Service Delivery Team