TGA Action ID RC-2026-RN-00532-1
Type of Product Medical Device
Product Name/Description SPECTRALIS with FLEX Module configuration

FLEX module product code: 86549
ARTG Number 222611
Action Type Product Correction
Action Level Hospital
Hazard Classification Class II
Reason for the Action There is a possibility that the castor wheels of the Flex stand are not properly attached. The wheels may come loose when the Flex Module is being pulled or pushed during movement or transportation, affecting the stability of the stand. As a result, the Flex stand may tip over if not correctly handled or supported by the operator. A falling Flex Module stand may cause impact or crushing injuries, particularly to the feet, legs, hands or arms of the operator or nearby persons.
Instructions for Customers/Users Customers should ensure that all castor wheels are properly aligned. Each wheel should stand upright and be in a straight position.
- If a wheel is upright and straight, it is correctly positioned
- If a wheel is tilted or leaning, it is not correctly positioned

Move the device only when necessary
If the wheels are tilted or leaning, do not move the stand anymore until the problem has been corrected.

In any case, please schedule a service appointment with the Heidelberg Engineering service partner as soon as possible so that the castor fastenings can be inspected and, if necessary, corrected
Action Commencement Date 15/07/2026
Responsible Entity Heidelberg Engineering Pty Ltd
Contact Information 1300 434 357 - Customer Service