| TGA Action ID |
RC-2026-RN-00532-1 |
| Type of Product |
Medical Device |
| Product Name/Description |
SPECTRALIS with FLEX Module configuration
FLEX module product code: 86549 |
| ARTG Number |
222611 |
| Action Type |
Product Correction |
| Action Level |
Hospital |
| Hazard Classification |
Class II |
| Reason for the Action |
There is a possibility that the castor wheels of the Flex stand are not properly attached. The wheels may come loose when the Flex Module is being pulled or pushed during movement or transportation, affecting the stability of the stand. As a result, the Flex stand may tip over if not correctly handled or supported by the operator. A falling Flex Module stand may cause impact or crushing injuries, particularly to the feet, legs, hands or arms of the operator or nearby persons. |
| Instructions for Customers/Users |
Customers should ensure that all castor wheels are properly aligned. Each wheel should stand upright and be in a straight position. - If a wheel is upright and straight, it is correctly positioned - If a wheel is tilted or leaning, it is not correctly positioned
Move the device only when necessary If the wheels are tilted or leaning, do not move the stand anymore until the problem has been corrected.
In any case, please schedule a service appointment with the Heidelberg Engineering service partner as soon as possible so that the castor fastenings can be inspected and, if necessary, corrected |
| Action Commencement Date |
15/07/2026 |
| Responsible Entity |
|
| Contact Information |
1300 434 357 - Customer Service |