Print version of this report
TGA Action ID RC-2026-RN-00531-1
Type of Product Medical Device
Product Name/Description CONSTELLATION ULTRAVIT (10K) and HYPERVIT (20K) probes, including probes supplied individually, in FMS (Fluid Management System) Procedure Paks, and in Custom Paks

ULTRAVIT (10K) Product Codes: 8065752448, 8065752450, 8065752452, 8065752435, 8065752437, 8065752439, 8065752417, 8065752415

HYPERVIT (20K) Product Codes: 8065000095, 8065830077, 8065753106, 8065000096, 8065753109, 8065830026, 8065830027

Manufactured before 8 Sep 2025
ARTG Number 145666, 319177, 370933
Action Type Recall / Product Alert
Action Level Hospital
Hazard Classification Class II
Reason for the Action There is potential for specific models of ULTRAVIT and HYPERVIT to unexpectedly fail to actuate and cut during use. These probes are intended for use with the CONSTELLATION Vision System and packs or kits.

In September 2025, Alcon issued a Product Alert (RC-2025-RN-00796-1) asking customers to reduce the actuation rate of affected probe products to no more than 5000 actuations per minute.

Alcon is now conducting a Recall for standalone ULTRAVIT and HYPERVIT probes and FMS (Fluid Management System) Procedure Paks containing these probes, and a Product Alert for Alcon Custom Paks containing these probes. Affected probes were manufactured before 8 Sep 2025.
Instructions for Customers/Users For ULTRAVIT probe, HYPERVIT probe, or FMS Pak products, customers are to:
1. Review inventory for any remaining affected probes manufactured before 2025-09-08.
2. Quarantine any unused affected product.
3. Contact Alcon Customer Service to organise a return and replacement of product.

For Alcon Custom Pak, customers are to:
1. Upon opening your Custom Pak for surgical use, review the manufacturing date of the probe products.
If the probe was manufactured before 2025-09-08, the probe should be used at a reduced actuation speed of no more than 5000 actuations per minute.

ULTRAVIT or HYPERVIT probes, and FMS (Fluid Management System) Procedure Pak or Custom Paks with these probes, that were manufactured on or after 2025-09-08, can be used for all actuation speeds as indicated in their associated labelling or Instructions for Use (IFU).
Action Commencement Date 14/07/2026
Responsible Entity Alcon Laboratories Australia Pty Ltd
Contact Information 1800 224 153 - Alcon’s Surgical Customer Services