| TGA Action ID |
RC-2026-RN-00520-1 |
| Type of Product |
Medical Device |
| Product Name/Description |
Philips Incisive CT System
Software Version: 4.5 & 5.0
Product Codes: 728143, 728144 |
| ARTG Number |
352131 |
| Action Type |
Product Correction |
| Action Level |
Hospital |
| Hazard Classification |
Class II |
| Reason for the Action |
Philips has identified a problem with Philips Incisive CT systems with software V4.5 & V5.0 due to a potential of failing to trigger clinical scan during Bolus Tracking.
This problem can only occur during a Bolus Tracking exam under the following specific conditions: • A previous scan had the auto-launch option enabled; • Image processing from a previous patient is complete during the current patient tracker scan; • The system pops up a message that requests operator confirmation to replace the images in the viewer (“Launch new series in the Review”); • The operator fails to timely confirm/close this pop-up message.
As a result of this event sequence, the threshold detection mechanism fails, and the Bolus Tracker scan continues beyond the expected threshold until the operator manually stops the scan or closes the pop-up message. This also leads to the clinical scan not triggering automatically.
This problem may result in the patient receiving additional radiation and delays in clinical procedure or diagnosis.
Philips has not received any reports of harm or injury associated with this problem as of May 2026. |
| Instructions for Customers/Users |
Customers are to: - Pay attention to the pop-up message during the Bolus Tracker exam and timely confirm/close the message. - Follow the Incisive CT IFU to keep monitoring the bolus values on the images during the Tracker scan to ensure the scan is proceeding as planned.
Philips will be in contact to schedule a time for a Philips Field Service Engineer (FSE) to visit and install a software solution to address this problem (Reference FCO72800856). The planned FCO release date is 31st Oct 2026. |
| Action Commencement Date |
20/07/2026 |
| Responsible Entity |
|
| Contact Information |
1800 251 400 - Philips Service Delivery Team |