TGA Action ID RC-2026-RN-00518-1
Type of Product Medical Device
Product Name/Description St Neonatal PICC Pack and St Paediatric Angiography Pack

St Neonatal PICC Pack
Product Code: 37-701
Batch Numbers: 061293026, 06729625, 06480325

St Paediatric Angiography Pack
Product Code: 38-768
Batch Numbers: 061415726, 06858825, 061044425, 061133526, 06362825, 06621525, 061567626
ARTG Number 163168
Action Type Recall
Action Level Hospital
Hazard Classification Class I
Reason for the Action The packs contain the item “Stopcock - 3 Way Lipid Resistant (Yellow) – item code: MMD IV110011” that is cracking or leaking when being used with propofol.

Cracking or leaking stopcocks could cause a sudden, unmonitored drop in the infusion rate or dosage. This could lead to unexpected patient awareness under anaesthesia or inadequate sedation. Cracks could create a breach in the closed sterile system, increasing the risk of catheter-related bloodstream infections.
Instructions for Customers/Users Customers are to:
- Inspect their stock and immediately quarantine any units of the affected batches to prevent further use.
- Return stock on hand with the completed response form to the sponsor.
- Contact their sale representative to arrange for a replacement/credit.

This action has been closed out on 25/08/2026
Action Commencement Date 9/07/2026
Responsible Entity Multigate Medical Products Pty Ltd
Contact Information 02 8718 2888 - Multigate Medical Products