Print version of this report
TGA Action ID RC-2026-RN-00511-1
Type of Product Medical Device
Product Name/Description BESDATA Single Use Scope Console / Image Processor

Product Code: MIP-1002

Serial Number: 251018011
ARTG Number 478085
Unique Device Identifier (UDI) (01)06971355352356(11)251018(17)301017(21)251018010
Action Type Recall
Action Level Hospital
Hazard Classification Class I
Reason for the Action The device label on the affected unit shows an incorrect model name and serial number.
The EndoScope WorkStation MIP-1003 SN# 251018010 was incorrectly labelled as MIP-1002 SN# 251018011.

The device’s casing did not feature a screw limiting function, appropriate labelling and warning statements regarding the internal lithium-ion battery.

A penetrated lithium battery may result in smoke, ignition or explosion.

This action does not affect any other Besdata products.

This action was undertaken prior to TGA notification.
Instructions for Customers/Users Customers are to:
- Complete the Customer Response form to confirm that the affected product has been discarded.

This action has been closed out on 03/08/2026
Action Commencement Date 9/07/2026
Responsible Entity ACN 684 152 994 Pty Ltd