TGA Action ID RC-2026-RN-00507-1
Type of Product Medical Device
Product Name/Description Flow Family Anesthesia Machines

Models: Flow-i C20, Flow-i C30, Flow-c

Part Numbers: 6677200, 6888520, 6677300, 6888530, 6887700
ARTG Number 180464
Unique Device Identifier (UDI) 07325710001349, 7325710010617, 07325710001363, 07325710010624, 07325710009765
Action Type Product Correction
Action Level Hospital
Hazard Classification Class II
Reason for the Action Maquet Critical Care AB has identified a problem that under certain conditions might prevent the device from performing as intended. In some cases, Flow family anaesthesia machines have not delivered anaesthetic agent when the vaporiser(s) were switched on.

These events have only been observed during the start-up phase of a case.
Instructions for Customers/Users Getinge requests that customers review the following:
1. Always confirm that anaesthetic agent is being delivered after activating the vaporiser(s), particularly at the start of a case.
2. According to the Instructions for Use, this indication signifies that concentration levels are below detectable limits, serving as confirmation that no agent is present.

** / gas analyzer has no information to display

3. Continuously monitor end-tidal anaesthetic agent concentration and patient vital signs to ensure adequate anaesthesia is maintained.
4. If delivery is not confirmed - end case and start up again. This measure will normally resolve the problem.

Getinge will be releasing a software update for software versions 4.10 and 4.11 to handle the no agent delivery problem.
Action Commencement Date 8/07/2026
Responsible Entity Getinge Australia Pty Ltd
Contact Information Quality.AUNZ@getinge.com