| TGA Action ID |
RC-2026-RN-00505-1 |
| Type of Product |
Medical Device |
| Product Name/Description |
Affera HexaGEN RF Generators and HexaPulse PF Generators
Software Versions 1.1.5 and 1.3.1
Model: AFR-00004 |
| ARTG Number |
467667 |
| Unique Device Identifier (UDI) |
00763000871840 and 00763000998318 |
| Action Type |
Product Correction |
| Action Level |
Hospital |
| Hazard Classification |
Class II |
| Reason for the Action |
As of March 2026, Medtronic has received a total of seventeen (17) commercial complaints related to software anomalies. Potential impacts associated with these anomalies include minor procedure delay, and improbable, theoretical risks of collateral tissue injury, transient heart block, or cancelled procedure. None of the theoretical risks have been observed, and no adverse events have been reported with these complaints. |
| Instructions for Customers/Users |
Customer Action Required: •Ensure access is provided for Medtronic to update the affected device. •Pass on this notice to all those who need to be aware and to any organisation where the potentially affected devices have been transferred. •Complete the customer response form and return it to rs.anzrecalls@medtronic.com. |
| Action Commencement Date |
3/07/2026 |
| Responsible Entity |
|
| Contact Information |
0417 406 225 - Alex Jackson (Medtronic) |