TGA Action ID RC-2026-RN-00502-1
Type of Product Medical Device
Product Name/Description Philips Azurion and Allura Systems

Affected Products: Allura Xper FD20, Allura Xper FD10, Allura Xper FD20, ALLURA XPER FD20 BIPLANE, Allura Xper FD20 OR Table, Allura Xper FD20/15, Azurion 7 B20, Azurion 7 M12, Azurion 7 M20, Azurion 5 M20

System Codes: 722006, 722010, 722012, 722013, 722026, 722028, 722035, 722058, 722068, 722078, 722079, 722223, 722224, 722226, 722228
ARTG Number 225815, 370112
Unique Device Identifier (UDI) N/A
Action Type Product Correction
Action Level Hospital
Hazard Classification Class II
Reason for the Action Philips has identified a potential safety problem with the Hard Disk Drives (HDDs) used in certain Philips Azurion and Allura systems.

In some cases, HDDs that are not aligned with the specifications stipulated by the legal manufacturer may have been installed during service activities.

Additionally, Philips has identified that in some Allura systems, HDDs compliant with specifications have been used within a single system, whereas specifications require identical HDD types.

These HDDs may exhibit reduced reliability over time, which may result in loss of imaging functionality. Additionally, in Allura systems, this may lead to loss of motorised movement or loss of data.

To date Philips has not received any reports of harm associated with this problem.
Instructions for Customers/Users Customers can continue to use their devices and follow the instructions included in the customer letter which state:

In case motorised movements are unavailable, the stand can be manually repositioned using the handgrips and brake controls located on both sides of the stand. For ceiling-mounted systems, manual longitudinal (1), transversal (2) and rotational (3) stand movements remain available.

Philips will inspect all affected systems and replace HDDs that are not according to the specifications, as well as correct configurations where different HDD types are used within a single system. Philips expects to start the implementation of this correction in September 2026.

A local Philips representative will contact you to schedule a visit for a Field Service Engineer to perform these activities.
Action Commencement Date 1/07/2026
Responsible Entity Philips Electronics Australia Ltd
Contact Information 1800 251 400 - Service Delivery Team