| TGA Action ID |
RC-2026-RN-00497-1 |
| Type of Product |
Medical Device |
| Product Name/Description |
The Mako 4 system running the Total Hip Arthroplasty (THA) 4.2 and THA 5.0 applications
TKA 2.1, THA 4.2, PKA 3.1 Install Disk Catalogue Number: 700002834690
THA 5.0, TKA 2.1, PKA 3.1 Install Disk Catalogue Number: 700003123526
TKA 3.0, THA 5.0, PKA 3.1 Install Disk Catalogue Number: 700002969752
TKA 3.0, THA 4.2, PKA 3.1 Install Disk Catalogue Number: 700003488242 |
| ARTG Number |
223106 |
| Action Type |
Product Correction |
| Action Level |
Hospital |
| Hazard Classification |
Class II |
| Reason for the Action |
There is the potential for the user to experience a discrepancy in the on-screen visualisation of bone resection and the actual amount of bone removed during the reaming process when using Total Hip Arthroplasty (THA) 4.2 and THA 5.0 applications on the Mako 4 Robotic Arm.
If the user observes excessive red and/or a black area in the acetabulum, this may lead to a surgical delay not expected to be more than 15 minutes.
There have been 23 complaints reported due to this problem, with no reports of adverse health consequences. |
| Instructions for Customers/Users |
If a customer observes excessive red and/or black area in the acetabulum, they are to follow the risk mitigations outlined in the customer letter. This includes using the CT view in the Reaming and Impaction pages to probe and visually confirm the probe tip to true resected surface, as well as the impacted cup's rim relative to the planned cup rim.
Stryker is currently investigating and developing a permanent solution. Software updates will be deployed through the regular service schedule of the Mako system. |
| Action Commencement Date |
1/07/2026 |
| Responsible Entity |
|
| Contact Information |
postmarketssp@stryker.com - Stryker Australia |