TGA Action ID RC-2026-RN-00495-1
Type of Product Medical Device
Product Name/Description Microbore Extension Sets and Lipid-Resistant Stop Cocks with Extension Sets (15 cm and 25 cm DEHP-Free)
Product Codes: IV050030, IV050140, IV050180, IV110002 & IV110003
All lots/batches

Stopcock - 3 Way Lipid Resistant (Yellow)
Product Code: IV110011
All lots/batches
ARTG Number 324222, 354371
Action Type Recall
Action Level Hospital
Hazard Classification Class I
Reason for the Action There have been reports of the products cracking or leaking when being used with propofol. This might result in the following problems:

• Cracks and leaks in the stopcock can cause a sudden, unmonitored drop in the infusion rate and unintended dosage drops. This can lead to unexpected patient awareness under anaesthesia or inadequate sedation.

• Cracks can create a breach in the closed sterile system increasing the risk of catheter-related bloodstream infections or sepsis.
Instructions for Customers/Users Customers are to:
- Inspect stock on hand and immediately quarantine all batches to prevent further use.
- Return stock on hand with the completed Customer Response Form to the sponsor.
- Contact the local sale representative to arrange for a credit.

This action has been closed out on 28/10/2026
Action Commencement Date 10/07/2026
Responsible Entity Multigate Medical Devices Pty Ltd
Contact Information 02 8718 2800 - Multigate Medical Devices