TGA Action ID RC-2026-RN-00473-1
Type of Product Medical Device
Product Name/Description NephroMax High Pressure Nephrostomy Balloon Catheters

Descriptions: NEPHROMAX 10-12/7/55, NEPHROMAX BALLOON KIT 30F W/17CM PTFE SH, NEPHROMAX 8-12/7/55, NEPHROMAX CLEAR 10-12/7/55,
NEPHROMAX 10-15/7/55, NEPHROMAX BALLOON KIT 24F W/17CM PTFE SH, NEPHROMAX BALLOON KIT 30F W/17CM PTFE CS, NEPHROMAX BALLOON KIT 30F W/20CM PTFE SH

Multiple material and lot numbers

Expiry Date: 30 June 2028 - 19 October 2028
ARTG Number 149646, 211738
Action Type Product Correction
Action Level Hospital
Hazard Classification Class II
Reason for the Action The renal sheath for certain NephroMax High Pressure Nephrostomy Balloon Catheters may be packaged in a reversed orientation.

If the sheath orientation is not corrected prior to use, this could increase resistance and result in difficulty advancing, withdrawing or positioning the catheter.
Instructions for Customers/Users Customers are to:
- Inspect the product before use, verify renal sheath orientation and correct if necessary.

If increased resistance is experienced during advancement or manipulation, stop, reassess, and correct the problem. Do not continue forceful manipulation.
Action Commencement Date 24/06/2026
Responsible Entity Boston Scientific Pty Ltd