| TGA Action ID |
RC-2026-RN-00461-1 |
| Type of Product |
Medical Device |
| Product Name/Description |
Integra Omni-Tract Surgical Retractor System
All lot and catalogue numbers |
| ARTG Number |
207255, 207257, 207258 |
| Action Type |
Product Correction |
| Action Level |
Hospital |
| Hazard Classification |
Class II |
| Reason for the Action |
Integra is issuing a correction to the Instructions for Use (IFU) for the Omni-Tract Surgical Retractor Systems. The IFU correction includes: - Update to the validated prevacuum steam sterilisation dry time parameters - Removal of rigid container use as an approved steam sterilisation configuration; and - Removal of gravity displacement steam sterilisation method.
If devices are not adequately dried, residual moisture can increase the potential for sterile barrier compromise leading to contamination transfer.
To date, there have been no reported complaints worldwide for wet products or patient harm.
UPDATE 12/08/2026: Additional sterilisation testing has been performed and Integra has updated the IFU to include a rigid container as an approved sterilisation parameter. |
| Instructions for Customers/Users |
Customers are advised to follow the updated sterilisation information in the customer letter until the updated IFU is made available.
If facilities have validated their own sterilisation process, it is in their discretion to follow their validated process, recognising that Integra LifeSciences has only validated the sterilisation process noted in the customer letter.
Customers should stop using the Gravity Sterilisation Method immediately. The Gravity Sterilisation Method for sterilising Omni-Tract devices has been determined as not effective.
UPDATE 12/08/26 - The updated IFU is now available. Customers are to replace their IFU copy with the updated version provided with the customer letter, or access a copy at https://labeling.integralife.com/eifu/
Integra has also undertaken additional sterilisation testing to include prevacuum sterilisation parameters when using a rigid container. |
| Action Commencement Date |
26/06/2026 |
| Responsible Entity |
|
| Contact Information |
1300 550 599 - Integra Neurosciences Customer Service |