TGA Action ID RC-2026-RN-00457-1
Type of Product Medical Device
Product Name/Description Sontek Swivel Y Airway Connector

Product Codes: SMI-1002 & SMI-1003

Expiry: July 2026 - Dec 2030
ARTG Number 309537
Action Type Product Correction
Action Level Hospital
Hazard Classification Class II
Reason for the Action Teleflex Medical Australia and the manufacturer Next Phase Medical Devices LLC is updating the Sontek Swivel Y Airway Connector Instructions for Use (IFU) to incorporate a new warning regarding catheter insertion. Users are advised not to force the catheter through the seal when excessive resistance is encountered as forced insertion may dislodge or damage the green seal.

In the event the green seal becomes dislodged the potential hazards include the inability to maintain positive-end expiratory pressure (PEEP) due to blockage, trapping of the dislodged seal in the airway, and inability to perform bronchoscopy due to blockage. In the worst-case scenario of blockage and inability to maintain PEEP, health risks include hypoxemia, decreased tidal volume, endotracheal tube dislodgement, bacteraemia, pneumonia including ventilator-assisted pneumonia, and hypoxia (secondary to alveolar collapse).

As of 01 July 2026, six complaints have been received from customers reporting a strong resistance and difficulty to push the scope though. No patient injuries have been reported at this time.
Instructions for Customers/Users Customers are to:
- Immediately check inventory for the affected product
- Place a copy of the Customer Letter with the IFU

The device can continue to be used safely as intended. Please follow all instructions and warnings contained in the amended IFU.

This action has been closed out on 15/10/2026
Action Commencement Date 6/07/2026
Responsible Entity Teleflex Medical Australia Pty Ltd
Contact Information 1300 360 226 - Teleflex Customer Service