TGA Action ID RC-2026-RN-00455-1
Type of Product Medical Device
Product Name/Description ARTIS icono Ceiling System

Product Code: 11328100
ARTG Number 273952
Action Type Product Correction
Action Level Hospital
Hazard Classification Class II
Reason for the Action Over time, some screws that keep the monitor attached may slowly become loose. If one or more screws at this connection loosen, a visible gap may appear between the monitor carrier and the X ray indicator light. If the condition is not detected in time or is ignored, additional screws may progressively loosen over time.

If operation continues while screws are loosening, internal cables routed within the arm may become damaged. This damage can result in loss of the displayed image.

In the worst-case scenario, complete loosening of the connection could occur. This may result in separation of the monitor carrier from the supporting arm and a downward drop.
Instructions for Customers/Users Customers are to:
- Regularly inspect for gaps between the monitor carrier and the ceiling-mounted arm. If observed, promptly contact your service organisation.
- Be attentive to any loss of displayed image, unusual noises, shaking, wobbling, or other unexpected behaviour during operation, as these may indicate loose screws. If observed, limit movements of the DCS contact your service organisation immediately for further instructions.

A service representative will contact customers to correct the screwed connection of the monitor ceiling support.

This action has been closed out on 19/11/2026
Action Commencement Date 16/06/2026
Responsible Entity Siemens Healthcare Pty Ltd
Contact Information 1800 310 300 - Siemens Customer Care Centre