TGA Action ID RC-2026-RN-00454-1
Type of Product Medical Device
Product Name/Description Hemo Dialysis Machine 4008S V10

Product code: M204001
ARTG Number 120309
Action Type Product Correction
Action Level Hospital
Hazard Classification Class II
Reason for the Action Sporadic flow alarms have been reported in 4008-HD devices. Investigation identified that the problem originates from minor technical variations in the motor manufacturer’s carbon brushes, which may wear faster than expected. Accelerated wear can lead to premature pump failure, as the rotor becomes insufficiently energised once the carbons are fully worn. This may result in failure of the degassing pump or flow pump, leading to loss of fluid pumping capability. No patient harm has been reported.

This market action was undertaken prior to TGA notification.
Instructions for Customers/Users Replacement for the affected motors was delivered in March 2026.

This action has been closed out on 16/07/2026
Action Commencement Date 23/06/2026
Responsible Entity Fresenius Medical Care Australia Pty Ltd
Contact Information 1800 011 464 - John Peeters, Fresenius Medical Care Australia