TGA Action ID RC-2026-RN-00452-1
Type of Product Medical Device
Product Name/Description Phantom Hindfoot TCC/TC Ball Tipped Guide Rod 3.0 x 800mm Sterile

Part Number: P31-955-1800-S

Lot Numbers: 501167202225, 501167302225, 501209514725, 501259627925, 501275832225
ARTG Number 289488
Unique Device Identifier (UDI) (01)00889795061577(17)300122(10)501167202225, (01)00889795061577(17)300122(10)501167302225, (01)00889795061577(17)300527(10)501209514725, (01)00889795061577(17)301006(10)501259627925, (01)00889795061577(17)301118(10)501275832225
Action Type Recall
Action Level Hospital
Hazard Classification Class II
Reason for the Action Two (2) complaints were received, involving four (4) units, regarding breaches of the sterile barrier on the pouches used to package the guide rod.

If the user does not identify the sterile barrier breach prior to surgery, this can cause the non-sterile device to be used and lead to infection. This infection can lead to further surgical intervention.
Instructions for Customers/Users Customers are to:
- Inspect their stock immediately and quarantine affected batches on hand to prevent further use.
- Return affected stock.
- Contact the manufacturer to arrange replacement stock or discuss alternative products.
Action Commencement Date 29/06/2026
Responsible Entity Actis Medical Pty Ltd
Contact Information (08) 7084 5277 - Actis Medical