Print version of this report
TGA Action ID RC-2026-RN-00447-1
Type of Product Medical Device
Product Name/Description CyberKnife Treatment Delivery Systems equipped with the Xchange Robotic Collimator Changer

Product Code: 0660000

Identifiers: C0345, C0447, C0596
ARTG Number 155887
Unique Device Identifier (UDI) C0345: +M6580534010/$$+7C0345/16D20180927U, C0447: M65806600000C04472017-11-30, C0596: 01008113760300161123120721C0596
Action Type Product Correction
Action Level Hospital
Hazard Classification Class II
Reason for the Action During a housing exchange operation, a collimator housing may not fully latch or release. This event triggers a system interlock while over the Xchange table that prevents motion from the operator console.

In this state, the collimator housing may fall from the robot while over the Xchange table. In normal use, patients should be removed from the treatment room following this system interlock.

While motion is prevented from the operator console, users may initiate robot motion using the in-room pendant. This motion increases the likelihood of the collimator housing falling.

No serious injuries have been reported.
Instructions for Customers/Users If an interlock occurs, customers are to:
(1) remove the patient from the treatment room before troubleshooting
(2) do not initiate treatment robot motion from the in-room control pendant
(3) contact Accuray Service (or your authorised service provider) to safely release the collimator housing.

An authorised service provider will contact customers to update all affected systems to software version 11.2.2.1.
Action Commencement Date 15/06/2026
Responsible Entity Emergo Asia Pacific Pty Ltd T/a Emergo Australia
Contact Information AUNZvigilance@ul.com - Emergo Australia