| Type of Product |
Medical Device |
| Product Name/Description |
CyberKnife Treatment Delivery Systems equipped with the Xchange Robotic Collimator Changer
Product Code: 0660000
Identifiers: C0345, C0447, C0596 |
| ARTG Number |
155887 |
| Unique Device Identifier (UDI) |
C0345: +M6580534010/$$+7C0345/16D20180927U, C0447: M65806600000C04472017-11-30, C0596: 01008113760300161123120721C0596 |
| Action Type |
Product Correction |
| Action Level |
Hospital |
| Hazard Classification |
Class II |
| Reason for the Action |
During a housing exchange operation, a collimator housing may not fully latch or release. This event triggers a system interlock while over the Xchange table that prevents motion from the operator console.
In this state, the collimator housing may fall from the robot while over the Xchange table. In normal use, patients should be removed from the treatment room following this system interlock.
While motion is prevented from the operator console, users may initiate robot motion using the in-room pendant. This motion increases the likelihood of the collimator housing falling.
No serious injuries have been reported. |
| Instructions for Customers/Users |
If an interlock occurs, customers are to: (1) remove the patient from the treatment room before troubleshooting (2) do not initiate treatment robot motion from the in-room control pendant (3) contact Accuray Service (or your authorised service provider) to safely release the collimator housing.
An authorised service provider will contact customers to update all affected systems to software version 11.2.2.1. |