Print version of this report
TGA Action ID RC-2026-RN-00442-1
Type of Product Medical Device
Product Name/Description Atellica CH Enzymatic Haemoglobin A1c (A1c_E) used with Atellica CH and Atellica CI systems. An in vitro diagnostic medical device (IVD).

Test Codes: A1c_E, A1c_H

Material Number: 11097536

All in-date lots
ARTG Number 176642
Unique Device Identifier (UDI) 00630414220505
Action Type Product Correction
Action Level Hospital
Hazard Classification Class II
Reason for the Action There is the potential for intermittent well-to-well bias affecting the A1c_E and A1c_H assays when used on the Atellica CH and Atellica CI systems.

This problem may result in:
· Falsely depressed patient results when lot calibration is performed using an affected well and patient testing is subsequently performed using an unaffected well.
· Falsely elevated patient results when lot calibration is performed using an unaffected well and patient testing is subsequently performed using an affected well.
Instructions for Customers/Users Customers are to perform QC on each reagent well used for all A1c_E/A1c_H testing.

Further information is provided in the customer letter regarding enabling QC and monitoring QC results.

In the event of QC failure:
o Unload the reagent pack and load a new pack.
o Repeat QC testing.
o After acceptable QC results have been obtained, repeat any held A1c_E/A1c_H samples since the failed QC event.

Report additional and/or discarded A1c_E/A1c_H tests or wells to Siemens Healthineers for credit.
Action Commencement Date 11/06/2026
Responsible Entity Siemens Healthcare Pty Ltd
Contact Information 1800 310 300 - Siemens Customer Care Centre