TGA Action ID RC-2026-RN-00431-1
Type of Product Medical Device
Product Name/Description Masimo RD rainbow 8ë SpCO sensors

MASIMO RD Rainbow Infant Adhesive Sensors (8ë SpCO, SpO2, SpMet)
Catalogue Number: 11996-000517
Lot: 23JAU

MASIMO LNCS-II Rainbow DCI 8ë SpCO, Adult Reusable Sensor
Catalogue Number: 11996-000519
Lots: 23JTA, 23HPT, 23HER, 23GJQ, 23JBN

MASIMO LNCS-II Rainbow DCI 8ë SpCO, Paediatric Reusable Sensor
Catalogue Number: 11996-000520
Lots: 23HNV, 23JBP
ARTG Number 201980
Action Type Recall
Action Level Hospital
Hazard Classification Class II
Reason for the Action Specific lots of Masimo Rainbow Sensors are incompatible with the LIFEPAK 15 V4 SpCO and SpMet functionalities.

Stryker received complaints indicating that the LIFEPAK 15 V4 displayed an error where “SpO2: Sensor does not support SpCO or SpMET.” The error results in preventing users from utilising SpCO and SpMet functionalities.

To date, there have been no adverse events reported to Stryker due to this problem.
Instructions for Customers/Users Customers are to:
1. Quarantine any affected product they have.
2. Contact their local Stryker representative to coordinate the replacement of any affected product.
Action Commencement Date 9/06/2026