TGA Action ID RC-2026-RN-00430-1
Type of Product Medical Device
Product Name/Description UHI-4 High Flow Insufflation Unit

SKU: N3829670, N3829660

All lot numbers
ARTG Number 211802
Action Type Product Correction
Action Level Hospital
Hazard Classification Class II
Reason for the Action Olympus identified an increase in UHI 4 front panel ‘blackout’ complaints, where the front panel display turns off unexpectedly, and air delivery (insufflation) ceases. An investigation found these blackouts (fault code E03) were primarily caused by the pressure sensors drifting positively from zero over time; when the pressure difference between the two sensors exceeds 7 mmHg, the UHI-4 generates the E03 fault, the front panel shuts off, and insufflation stops.
Instructions for Customers/Users Olympus is working to implement a long-term solution for front panel blackout. In the interim, the UHI-4 pressure sensors must receive an offset adjustment to reset the sensor to zero because of sensor drift. This adjustment can be completed onsite and does not require returning the unit to Olympus. An Olympus field service representative will contact facilities to schedule a mutually convenient time to perform the offset adjustment.

If the UHI-4 displays a front panel blackout (fault code E03) during use, either reboot the unit or switch to a backup insufflator. To reboot the UHI-4, turn the unit OFF and ON again. If during reboot, the caution continuously sounds, the reboot did not work, switch to the backup insufflator and contact Olympus.
Action Commencement Date 22/06/2026
Responsible Entity Olympus Australia Pty Ltd
Contact Information 1300 132 992 - Olympus Customer Operations