| TGA Action ID |
RC-2026-RN-00419-1 |
| Type of Product |
Medical Device |
| Product Name/Description |
Dolphin Imaging Software
Version: 13.0.10.34 |
| ARTG Number |
287656 |
| Action Type |
Product Correction |
| Action Level |
Retail |
| Hazard Classification |
Class II |
| Reason for the Action |
Following initial deployment activities of version 13.0.10.34, problems were identified that may affect how normative reference datasets are applied during orthodontic and cephalometric analysis. These challenges originate from patient archive-based export/import processes and system display logic.
• Problem 1: Measurements – Normative values displayed in the Measurements table may reflect an incorrect patient gender. • Problem 2: Archive –During archive-based data migration from Version 12.0 to 13.0.10.34, the patient gender field may change under specific conditions. This may result in the application of inappropriate gender-based normative datasets following migration.
Normative datasets are used as reference baselines to contextualise patient measurements. If an incorrect normative dataset is applied, analytical outputs may be misinterpreted, such that results may appear abnormal when they are normal or appear normal when they are abnormal. No complaints or adverse events have been received to date. |
| Instructions for Customers/Users |
Users should manually verify patient demographics and normative dataset selection prior to analysis. Any inconsistencies should be corrected before results are used for clinical interpretation.
Deployment of Dolphin Imaging 13.0.10.34 has been temporarily halted. A software update addressing these problems is currently in development, and customers will be notified once it is available, including instructions for implementation. |
| Action Commencement Date |
3/06/2026 |
| Responsible Entity |
|
| Contact Information |
(02) 9006 1662 - Emergo Australia |