TGA Action ID RC-2026-RN-00411-1
Type of Product Medical Device
Product Name/Description Sigma-Virocult 12x80 Vial with 1.5ml Liquid Virus Transport Medium. An in vitro diagnostic medical device (IVD).

Product Code: MW951T15

Batch Number: 144199

Expiry Date: 31/10/2027
ARTG Number 373837
Action Type Recall
Action Level Hospital
Hazard Classification Class II
Reason for the Action A number of transport medium vials have been reported to have microorganism contamination in the medium. The presence of the microorganism contaminant may result in assay interference and an erroneous result. The test control may also be affected and result in an invalid result.

The erroneous or invalid results may lead to incorrect and/or delayed patient treatment.

No other batches of Sigma-Virocult 12x80 Vial with 1.5ml Liquid Virus Transport Medium are affected by this problem.
Instructions for Customers/Users Customers are to inspect stock and quarantine units from batch 144199 to prevent futher use. Affected stock should be returned per the instructions in the Customer Letter.

Replacement stock of an alternative batch number is available and will be shipped to your organisation. Our customer service team will be in contact to arrange shipment as soon as possible

This action has been closed out on 08/07/2026
Action Commencement Date 2/06/2026
Responsible Entity Southern Cross Diagnostics Pty Ltd