TGA Action ID RC-2026-RN-00403-1
Type of Product Medical Device
Product Name/Description Iris Retractor MST/ MAL-1001 – Malyugin Ring 2.0 6.25mm

Lot Number: 206602

Expiry Date: 30/05/2028
ARTG Number 279045
Unique Device Identifier (UDI) 10844685006127
Action Type Recall
Action Level Hospital
Hazard Classification Class II
Reason for the Action There was one report of a Malyugin Ring exhibiting deformation during use, including bent struts and distorted scrolls. This prevented full engagement of the scrolls with the iris edge, resulting in inadequate fixation and device instability. The partially engaged, misshapen scrolls generated frictional forces on the iris during manipulation, leading to damage to the patient’s eye.
Instructions for Customers/Users Customers are to:
- Discard any faulty device identified and use an alternative device to mitigate the risk of harm if the device appears or performs abnormally.
- Inspect affected stock immediately and follow the instructions provided in the customer letter to reconcile this recall.

This action has been closed out on 04/08/2026
Action Commencement Date 27/05/2026
Responsible Entity IQ Medical
Contact Information (08) 6275 7820 - Lucinda Brooks, IQ Medical