| TGA Action ID |
RC-2026-RN-00396-1 |
| Type of Product |
Medical Device |
| Product Name/Description |
VITROS Chemistry Products SALI Slides. An in vitro diagnostic medical device (IVD)
Affected Generations (GENs): 04 and above Catalogue Number: 131 4343
Impacted Systems: VITROS XT 3400 Chemistry System, VITROS XT 7600 Integrated System, VITROS XT 7600 Integrated System - Certified Catalogue Numbers: 684 4458, 684 4461, 627 2222
Affected Assay Data Disk Data Release Versions (ADD DRV): Below 6375 |
| ARTG Number |
232245 |
| Action Type |
Product Correction |
| Action Level |
Hospital |
| Hazard Classification |
Class II |
| Reason for the Action |
QuidelOrtho has identified a problem involving Condition Code TH4-63J that may cause calibration failures and/or an inability to generate patient results when using SALI Slides (GEN 04 and above) on VITROS XT 3400 and XT 7600 systems.
The problem occurs when the reflectometer detects image variations exceeding assay parameters on the Assay Data Disk (ADD). To address this, the assay parameter was adjusted starting with ADD DRV 6375.
This problem does not cause erroneous results. If patient results were generated, the results are valid. |
| Instructions for Customers/Users |
Customers are to: - Install ADD DRV 6375 (or above). |
| Action Commencement Date |
22/05/2026 |
| Responsible Entity |
|
| Contact Information |
1800 032 359 - Technical Services Center, QuidelOrtho |