| TGA Action ID |
RC-2026-RN-00394-1 |
| Type of Product |
Medical Device |
| Product Name/Description |
GORE TAG Thoracic Branch Endoprosthesis Side Branch Component
Catalogue Numbers: TSB080806X, TSB081006X, TSB081206X, TSB081506X, TSB121506X, TSB081706X, TSB121706X, TSB122006X
Manufactured between 05 August 2025 and 20 April 2026 |
| ARTG Number |
504789 |
| Unique Device Identifier (UDI) |
073313230255602136 |
| Action Type |
Product Alert |
| Action Level |
Hospital |
| Hazard Classification |
Class I |
| Reason for the Action |
Reports have been received of catheter separation involving the GORE TAG Thoracic Branch Endoprosthesis (TBE) Side Branch (SB) Component during clinical use. The separation occurs at the interface between the main catheter shaft and an over-moulded transition near where the trailing end of the device is positioned on the catheter, with all reported events occurring after deployment of the Side Branch component.
In the five reported events, contributing factors included higher than typical force being applied during delivery or removal of the delivery system and, in some cases, catheter rotation. |
| Instructions for Customers/Users |
Physicians are to: - Be reminded of the Instructions for Use (IFU), including the information, warnings, and additional considerations outlined in the Customer Letter. - Consider this information in patient-specific risk–benefit decisions.
This action has been closed out on 14/10/2026 |
| Action Commencement Date |
26/05/2026 |
| Responsible Entity |
|
| Contact Information |
(02) 9473 6800 - W. L. Gore & Associates |