| TGA Action ID |
RC-2026-RN-00384-1 |
| Type of Product |
Medical Device |
| Product Name/Description |
LIFECODES LifeScreen XP Kit. An in vitro diagnostic medical device (IVD).
Product Number: 628220
Lot Number: 3016386
Expiration Date: 15 SEP 2026 |
| ARTG Number |
217870 |
| Unique Device Identifier (UDI) |
10888234500841 |
| Action Type |
Product Alert |
| Action Level |
Hospital |
| Hazard Classification |
Class III |
| Reason for the Action |
The LXP Negative Control Serum (LXPNC) for lot 3016386 may, in some instances, produce unexpected positive results, which can lead to assays being classified as invalid according to the Instructions for Use (IFU) acceptance criteria.
Patient risk is considered low because results from assays that do not meet validity criteria are not reportable. |
| Instructions for Customers/Users |
Customers may continue to use the affected kit lot number in accordance with the acceptance criteria labelled in the IFU for LIFECODES LifeScreen XP. LXP Negative Control Serum (LXPNC) with negative assignments are considered valid.
Contact Werfen Customer Support if an increase in rate of invalid tests is observed.
There is no general requirement to retest samples that met acceptance criteria labelled in the IFU for LIFECODES LifeScreen XP Kit. However, laboratories should consider the points outlined in the Customer Letter.
This action has been finalised on 21/9/2026. |
| Action Commencement Date |
27/05/2026 |
| Responsible Entity |
|
| Contact Information |
1800 788 751 - Werfen Customer Support |