| TGA Action ID |
RC-2026-RN-00376-1 |
| Type of Product |
Medical Device |
| Product Name/Description |
Philips Azurion Systems
Name: Azurion 7 B12 Code: 722067, 722235
Name: Azurion 7 B20 Code: 722068, 722226, 722236 Name: Azurion 7 M12 Code: 722078, 722223, 722233
Name: Azurion 7 M20 Code: 722079, 722224, 722234
Name: Azurion 5 M12 Code: 722227
Name: Azurion 5 M20 Code: 722228
Systems already configured with an auxiliary pan handle are subject to a Product Alert. Systems NOT configured with an auxiliary pan handle are subject to a Product Correction. |
| ARTG Number |
225815 |
| Action Type |
Product Correction / Product Alert |
| Action Level |
Hospital |
| Hazard Classification |
Class II |
| Reason for the Action |
Philips has identified a potential safety problem with Philips Azurion systems. The Float Tabletop control is located on the control module at the table-side. When the control is actuated, the table brakes disengage, allowing adjustment of the longitudinal or transverse position of the table.
Over time, mechanical wear may lead to detachment, cracking, or breakage of the Float Tabletop control, which may lead to reduced availability or unavailability of longitudinal and transverse tabletop movement, as well as the formation of sharp edges. |
| Instructions for Customers/Users |
For all systems: - Avoid contact with sharp edges in the event that the Float Tabletop control is damaged. - During the Daily System Check indicated in the IFU, verify the system functionality including testing the table. - Should you experience detachment, cracking, or breakage of the Float Tabletop control or reduced or unavailable longitudinal and transversal tabletop movement, please report this to Philips through your local Philips representative.
For Azurion systems NOT configured with an optional auxiliary pan handle: - If the required position cannot be achieved using the Float Tabletop control for table movement, use alternative system movements (e.g., stand movement) where clinically feasible to adjust the relative position between the system and the patient. - Philips will install an auxiliary pan handle. A Philips representative will contact you to schedule the implementation of this correction. Philips expects to begin implementation of the correction by July 2026. |
| Action Commencement Date |
9/06/2026 |
| Responsible Entity |
|
| Contact Information |
1800 251 400 - Philips Service Delivery Team |