TGA Action ID RC-2026-RN-00373-1
Type of Product Medical Device
Product Name/Description Apollo - Robot Enclosure Body Assembly

Part Number: TKR.OB.1000

Housing ID: 1741 / 1769 / 2102
ARTG Number 450027
Action Type Product Correction
Action Level Hospital
Hazard Classification Class III
Reason for the Action The manufacturer, Corin Limited, has identified that the Axial Screw (4145-2005) of the Apollo Robot/OMNIBot Motor Mount Assembly (NS-41020), a sub-component of TKR.OB.1000 (Apollo Robot Enclosure Body Assembly) was manufactured out of specification and are all at risk of shearing failure when tightened.

The risks of harm to the patient are not high in the failure modes observed, with the worst-case impact expected to be a potential for small delay to the normal surgical time if a failure was detected late in the procedure.
Instructions for Customers/Users Customers are to quarantine and return the affected stock on hand to Corin Australia. The same device will be returned to the customer once corrected.

As the affected parts are sub-components of TKR.OB.1000 (Apollo Robot Enclosure Body Assembly), the whole Apollo Robot Enclosure Body Assembly is to be returned to replace the at-risk 4145-2005 screws for re-work.

This action has been closed out on 24/08/2026
Action Commencement Date 21/05/2026
Responsible Entity Corin Australia Pty Ltd
Contact Information (02) 9497 7400 - Corin Australia