| TGA Action ID |
RC-2026-RN-00373-1 |
| Type of Product |
Medical Device |
| Product Name/Description |
Apollo - Robot Enclosure Body Assembly
Part Number: TKR.OB.1000
Housing ID: 1741 / 1769 / 2102 |
| ARTG Number |
450027 |
| Action Type |
Product Correction |
| Action Level |
Hospital |
| Hazard Classification |
Class III |
| Reason for the Action |
The manufacturer, Corin Limited, has identified that the Axial Screw (4145-2005) of the Apollo Robot/OMNIBot Motor Mount Assembly (NS-41020), a sub-component of TKR.OB.1000 (Apollo Robot Enclosure Body Assembly) was manufactured out of specification and are all at risk of shearing failure when tightened.
The risks of harm to the patient are not high in the failure modes observed, with the worst-case impact expected to be a potential for small delay to the normal surgical time if a failure was detected late in the procedure. |
| Instructions for Customers/Users |
Customers are to quarantine and return the affected stock on hand to Corin Australia. The same device will be returned to the customer once corrected.
As the affected parts are sub-components of TKR.OB.1000 (Apollo Robot Enclosure Body Assembly), the whole Apollo Robot Enclosure Body Assembly is to be returned to replace the at-risk 4145-2005 screws for re-work.
This action has been closed out on 24/08/2026 |
| Action Commencement Date |
21/05/2026 |
| Responsible Entity |
|
| Contact Information |
(02) 9497 7400 - Corin Australia |