TGA Action ID RC-2026-RN-00369-1
Type of Product Medical Device
Product Name/Description CASE v7.0 Cardiac Testing Systems

Product Code: SU3

GTIN: 00195278617644
ARTG Number 362792
Action Type Product Correction
Action Level Hospital
Hazard Classification Class II
Reason for the Action GE HealthCare has become aware of the potential for electrical sparking at the printer head of CASE v7 systems. In the unlikely situation that the Electrostatic Discharge (ESD) printer brush becomes dislodged, electrical sparking at the printer head can result in charring or discoloration of the thermal printer paper. In rare cases, ignition of the paper could result in serious thermal injury.

There have been no injuries reported to GE HealthCare as a result of this problem.
Instructions for Customers/Users Customers are to perform a visual inspection of the printer assembly to verify proper positioning of the ESD brush as outlined in the Customer Letter.

If the brush is properly positioned and aligned, continue to use the device.
If the brush appears misaligned or displaced, stop using the device and contact GE HealthCare Service.

GE HealthCare will install a new ESD static brush and conductive shield if the static brush is installed incorrectly or damaged.
Action Commencement Date 25/05/2026
Responsible Entity GE Healthcare Australia Pty Ltd
Contact Information 1800 659 465 - GE HealthCare Service